38
Attaching the Pulse Oximetry Module
Should you need to attach or remove a pulse oximetry module, the following instructions
give an overview of its attachment to the BP module. The pulse oximetry module attaches to
the bottom of the BP module.
1.
Using the power button on the right side of the BP module, ensure
that the
SunTech
247
is off.
2.
Remove the cover plate from the bottom of the main BP module.
3.
Slide the pulse oximetry module along the guides on the bottom of the
main BP module from front to back until it snaps into place. All the
segments of the pulse oximetry module display will light when the
modules have been connected correctly.
4.
Turn the device on using the power button on the right side of the BP module. At
the end of the start-up sequence, the pulse oximetry module display will blank
except for the ―%SpO2‖ symbol.
5.
Turn the device off. Connect the sensor or extension cable to the connector on the
module. Secure the cable to your device using the retention clip.
6.
Turn the device on. All display segments light up for three to five seconds. A short
beep indicates that the
SunTech 247
is ready.
Note: If the pulse oximetry module does not appear to be working properly, cycle the power
several times using the power button on the right side of the BP module. This will
―synchronize‖ all the modules. The modules are synchronized when, after you turn the
power on, all segments on all module displays light simultaneously for 3 to 5 seconds,
followed by a short beep, and all displays go to their ―ready‖ state (BP: battery icon and
middle segments of the systolic value are lit; Temperature: appropriate temperature units icon
is lit; Pulse Oximetry: the ―%SpO2‖ icon is lit).