MEASURING OXYGEN SATURATION WITH THE
SUNTECH
247
The
SunTech 247
optional pulse oximeter modules measure functional oxygen saturation
ranging from 40% to 100%. A signal strength display assists the clinician in the proper
placement of the sensor.
The
SunTech 247
is a spot check device and is not used for patient
monitoring. Therefore, there are no SpO
2
alarms.
Steps for measuring functional oxygen saturation:
Prepping the patient and affixing the sensor
Taking a reading
Prepping the Patient
Selecting the Right Sensor
Your choice of sensor is affected by many factors including:
Patient‘s body weight
Patient activity
Infection control concerns
For most patients greater than 30kg, use an adult sensor; for patients 10-
50kg, a pediatric sensor may provide better fit.
Disposable sensors can provide a more secure connection to the patient.
Protecting Your Pulse Oximetry Sensors
In addition to the safety instructions for your
SunTech 247
, here are some additional tips on
caring for the sensors of the optional pulse oximetry modules:
To prevent damage, do not autoclave or immerse the sensor in liquid.
For peak performance and accurate measurements, do not expose the sensors to
excessive ambient light, electromagnetic interference, dysfunctional hemoglobin, low
perfusion, intravascular dyes, finger nail polish and long or artificial finger nails.