Buttons on the main BP module
Button Functionality for the BP Module
Button
Device Status
Action
Result
Automatic
Idle
Select for less than 2
seconds
Start an automatic BP
measurement.
Select for 2 to 5
seconds
Redisplay last measurement values
for all modules.
Select for more than 5
seconds
Clear last measurement values for
all modules.
Taking an
automatic or
SphygMode
BP
Select
Aborts the BP measurement in
progress.
In Calibration
Check mode
Select
Device exits Calibration Check
mode and is ready to take
measurements.
SphygMode
Manual
Idle
Select
Inflates the cuff as long as the
button is selected.
Taking a
SphygMode
BP
Select
Re-inflates the cuff as long as the
button is selected.
Taking an
automatic BP
Select
Aborts the BP measurement in
progress.
In Calibration
Check mode
Select
Device exits Calibration Check
mode and is ready to take
measurements.
Power
Power off
Select
Turns on the device.
Power on
Select
Turns off the device.
Aut
Power
Power off
Hold the Automatic
button down while
selecting the Power
button.
Device enters the Calibration
Check mode.
SphygMode
Power
Power off
Hold the
SphygMode
button down while
selecting the Power
button.
Device enables/disables MAP