45
supplements this with the ability to take measurements as you would if you were using a
mechanical sphygmomanometer.
Can I upgrade my current version of
SunTech 247
at a later date?
To upgrade your
SunTech 247
device, please review the list of accessories on page 61 or on
our website. Contact your local distributor for details.
Could I use the
SunTech 247
to measure blood pressure during a stress test?
Although your
SunTech 247
is a robust device that has been manufactured with motion
tolerance, it is not intended for use during stress testing. Please use an alternate device like
the SunTech
Tango
+ for treadmill or ergometer stress.
Should I wait between temperature measurements?
Yes. Accurate temperature measurement requires the probe to be at normal room
temperature. After taking a measurement, wait for the probe to return to room temperature
or wipe the probe with an alcohol wipe before taking a subsequent measurement.
Do I need to calibrate the temperature or pulse oximetry modules?
For calibration or service on the
SunTech 247
temperature and pulse oximeter modules,
contact an authorized service center on page 46 - 46.
How accurately does the
SunTech 247
temperature module measure temperature?
The
SunTech 247
temperature module is accurate to +/- 0.2
F (+/-0.1
C).
What is the accuracy of the
SunTech 247
pulse oximetry module?
The
SunTech 247
pulse oximetry module is +/- 2% in the 70 to 100% range for no motion and
normal perfusion. For motion or low perfusion, the accuracy is +/- 3%.