Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity
should be at least 30 %.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Mesh Nebulizer with
Rechargeable Battery requires
continued operation during
power mains interruptions, it is
recommended that the Mesh
Nebulizer with Rechargeable
Battery be powered from an
uninterruptible power supply or
a battery.
Guidance and manufacturer's declaration – electromagnetic immunity – for all
EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration –
electromagnetic immunity
The Mesh Nebulizer with Rechargeable Battery is intended for use in the
electromagnetic environment specified below.
The customer or the user of the Mesh Nebulizer with Rechargeable Battery
should assure that it is used in such an environment.
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrostatic
transient / burst
IEC 61000-4-4
Voltage dips,short
interruptions and
voltage variations
on power supply
input lines EN
61000-4-11
Surge
IEC 61000-4-5
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
± 2 kV for power
supply lines
± 1 kV differential
mode
0% UT;0.5 cycle
At
0°/45°/90°/135°/
180°/225°/270°and
315°
0 % UT;1 cycle and
70 % UT;25/30
cycles
Single phase: at 0°
0 % UT; 250/300
cycle
N/A
N/A
Recommended separation
distance
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
NOTE
:
UT is the a. c. mains voltage prior to application of the test level.
30 A/m
30 A/m
N/A
N/A
Immunity test
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Mesh
Nebulizer with Rechargeable
Battery ,including cables,
than the recommended
separation distance
calculated from the equation
applicable to the frequency
of the transmitter
Guidance and manufacturer´s declaration – electromagnetic immunity –for
EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Guidance and manufacturer´s declaration –
electromagnetic immunity
The Mesh Nebulizer with Rechargeable Battery is intended for use in the electromag-
netic environment specified below. The customer or the user of the Mesh Nebulizer
with Rechargeable Battery should assure that it is used in such an environment.
Conducted RF
IEC 61000-4-6
EN 60601
test level
Compliance
level
Electromagnetic
environment - guidance
3V rms
150kHz to 80MHz
6V in ISM bands
between 0
.
15MHz
and 80 MHz
Immunity test
28
29
d=[3.5/V1]