Picture 13 Aerosol Particle Equivalent Volume
(1) It was tested by Pulmicort under the condition of temperature 24±2
℃
,
humidity 45-75%R.H, Pressure: from 86 kpa to 106 kpa.
(2) MMAD (µm): 2.2µm ±25%
Respirable fraction (%, 0.5-5 µm) : ≥ 80%
Aerosol output rate (ml/min) : 0.24 ± 0.02 ml/min
The Mesh Nebulizer with Rechargeable Battery has been tested and found to comply
with the limits for medical devices in IEC 60601-1-2. These limits are designed to provide
reasonable protection against harmful interference in a typical medical installation.
If other devices cause harmful interference to this device, the user is encouraged to try
to correct the interference by one or more of the following measures:
Reorient or relocate this device.
Increase the separation between the interference equipment.
Connect the equipment into an outlet on a circuit different from that to which the
other device(s) are connected.
Consult the manufacturer for help.
1. Electromagnetic Compatibility
Do not connect this equipment to other equipment. Consult your physician for users
wearing a heart pacemaker.
Do not use the device in the direct vicinity of short-wave or microwave equipment.
Warning:
Use of this equipment adjacent to
or stacked with other equipment should be avoided
because it could result in improper operation. If such
use is necessary, this equipment and the other
equipment should be observed to verify that they
are operating normally
.
Warning:
Use of accessories, transducers
and cables other than those specified or provided by
the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and
result in improper operation.
Warning:
Portable RF
communications equipment
(including peripherals such as
antenna cables and external
antennas) should be used no
closer than 30 cm (12 inches)
to any part of the [ME
EQUIPMENT or ME SYSTEM],
including cables specified by
the manufacturer. Otherwise,
degradation of the
performance of this
equipment could result.
Guarantee Time
Main Unit 1 Year (Medicine Cup is not included)
Operating Temperature
Environmental temp:5
℃
~40
℃
,
Relative humidity : 15~90% RH, not in Coagulation situation
Air Pressure: (70.0~106.0)kPa
Storage and transport
Temperature
Accessory
Environmental temperature: -25
℃
~70
℃
Relative Humidity: ≤90%R.H.
Air Pressure: (70.0~106.0)kPa
Mouthpiece, adult Mask, child Mask, Adapter,
Instruction Manual
24
25
Chapter 9 Electromagnetic Compatibility
NOTE !
Aerosol particle equivalent volume (Picture 13)