2. Disinfecting
1. Medicine Cup
After cleaning, Immerse the Medicine Cup by using the following high-level disinfection
:
Disinfection Agent: Cidex OPA
Concentration: 0.55%
Contact Type: Soak
Time: 5 min
Temperature: 20 °C
Then rinse cup and dry with medical gauze.
Picture 11 cleaning
water
(1) It is recommeded to pour out any residual liquid after each use to prevent any
mesh obstruction.
(2) In order to prevent thermal deformation,boiling water is not allowed.
(3) DO NOT wipe the Medicine Cup with cotton paper or other cloth. This may cause
residual particles to block the mesh enclosure.
(4) DO NOT touch mesh with fingers,swabs,or any objects to avoid damage.
NOTE !
1.4 Wipe the chambers of the Medicine Cup with a lint-free towel moistened with 70%
ethyl alcohol at least 3 times and air dry; Then re-add purified water (3-6 ml).
1.5 Press and hold the Power button for 3 seconds WHILE THE DEVICE IS OFF, this will turn
on the light and alternate between green and blue colors. Release the button
allowing the device to enter into cleaning mode for 2 minutes. The purified water is
"nebulized" to clear up any residual liquid on the mesh of the Medicine Cup.
wipe
wipe
keep pushing
Picture 12
2. Main Unit
2.1 After use, gently wipe the Main Unit and Electrodes (connecting the Main Unit and
Medicine Cup) with a dry clean medical gauze to ensure an effective connectivity
between them.
2.2 If there is still debris, clean the Main Unit and Electrodes with a medical gauze
soaked with 70% ethyl alcohol, and wipe off the excess liquid to make the device
work normally.
DO NOT wipe the device with volatile liquid (e.g. benzene, gasoline, diluent).
DO NOT touch the electrodes outside the device and the Medicine Cup with
cotton swab or other things, which would damage the electrodes.
DO NOT immerse the main device in any liquid to avoid damage.
NOTE !