4. Storage and transport

1. Storage and transport Conditions:

(1) Temperature: -13°F — 158°F (-25°C — 70°C)
(2) Relative Humidity: ≤90% R.H.
(3) Air Pressure: 70.0 ~ 106.0 kPa;
(4) Other requirements: No direct sunlight, Ventilation, No corrosive gas, 

     No heating device, or Open fire.

3. Cleaning and Disinfecting of Mouthpiece and Masks

The masks and mouthpiece provided are disposable. If you plan to use masks and 
mouthpieces provided by other manufacturers, please follow the instructions for 
cleaning and disinfection provided by them.

(1) To prevent disinfecting agent from touching skin, please wear rubber gloves.
(2) Do not wipe the mesh with tissue paper or other cloth, otherwise the mesh will 

not nebulize normally.

(3) Do not wipe or touch the mesh with your hands or any other sharp objects.
(4) Do not use household bleach to clean the devcie.

NOTE !

5. Maintenance

6. To protect the environment

(3) If the device has been stored at low or high temperature, the device needs    more 
than half an hour to adjust, until it is ready for intended use.
Validity: When nebulizing 0.9% sodium chloride solution for 2 times a day, and 15  
minutes for each time, 3 years for main unit, 3 months for Medicine Cup. For the battery, 
the device can be used for approximately 8 days after charging if operating for 30 
minutes a day. Lithium battery has 300 charge-discharge cycle. Masks and mouthpiece 
are single use only, please do not use repeatedly.

The battery is not replaceable. Do not discard the device, including its Accessories 
(mouthpiece and masks), in household trash. Follow local regulations to determine the 
proper method of disposal.

1.Working Conditions

(1) Environment Temperature: 41°F — 104°F (5°C — 40°C)
(2) Relative Humidity: 15% ~ 90% R.H No condensation; 
(3) Air Pressure: 70.0 ~ 106.0 kPa;
(4) Power: DC 3.7V lithium battery. 

2. Maintenance Instructions

(1) Operate the device in normal work condition;
(2) Don't let the Nebulizer and its Accessories contact corrosive liquids or gases.
(3) Refer to Chapter 7 for solution or contact the company or purchase dealer when 

the device cannot work.

(4) Please clean and disinfect the nebulizer as stated in this Chapter.
(5) Please charge the device before use if you haven't used the device for more than 

six months.

2. Storage Instructions:

(1) Keep the device out of the reach of unsupervised children.
(2) Clean and disinfect the device after using. Assemble and store the device in 
the storage bag or in a clean environment.

Содержание NEB401

Страница 1: ...uction Manual Mesh Nebulizer with Rechargeable Battery NEB401 Rx Only Manufactured for Sunset Healthcare Solutions 180 N Michigan Ave Ste 2000 Chicago IL 60601 www sunsethcs com NEB401 IFU V1 01 2020...

Страница 2: ...nt the device may be used with pediatric 4 years of age and adult patients in the home hospital and sub acute care settings It is not intended for use with Pentamidine Table of Contents Chapter 1 Befo...

Страница 3: ...ooting Chapter 7 Customer Service Electromagnetic Compatibility Notes on FCC certification Chapter 9 Electromagnetic Compatibility Symbol Glossary Chapter 10 Symbol Glossary Cleaning Disinfecting Clea...

Страница 4: ...with pediatric 4 years of age and adult patients in the home hospital and sub acute care settings It is not intended for use with Pentamidine 1 What is the Mesh Nebulizer with Rechargeable Battery 2...

Страница 5: ...nvironment Warning For type or dose of medicine please follow the instructions of your physician and or licensed healthcare practitioner Important Safety Information Read Before Use The Mesh Nebulizer...

Страница 6: ...on in the treatment area such as sores herpes simplex or open wounds If you are not sure the device is safe for you or not please consult your doctor PRECAUTIONS Take the following precautions to avoi...

Страница 7: ...e end of USB cord to an adapter or other USB port such as USB port on a computer This nebulizer is powered with a built in lithium battery The battery can be charged with the USB cord 1 Power Supply C...

Страница 8: ...ed in the right place the device may not be able to operate normally NOTE Picture 5 Mask Assembly and Mouthpiece Assembly 4 Usage of Nebulizer Nebulizer Accessories are categorized as mouthpiece adult...

Страница 9: ...hile nebulizing 1 If this nebulizer is being used to treat a life threatening condition a backup device is necessary 2 The nebulizer may be unable to be used due to power failure or mechanical malfunc...

Страница 10: ...n off the power The Power Indicator Green Light will go out 1 Cleaning Chapter 5 Maintenance of the Nebulizer Be sure to turn off the device After each use first discard all residual medicine and remo...

Страница 11: ...r 3 6 ml 1 5 Press and hold the Power button for 3 seconds WHILE THE DEVICE IS OFF this will turn on the light and alternate between green and blue colors Release the button allowing the device to ent...

Страница 12: ...utes for each time 3 years for main unit 3 months for Medicine Cup For the battery the device can be used for approximately 8 days after charging if operating for 30 minutes a day Lithium battery has...

Страница 13: ...se under guidance of doctor s prescription The device is on but weak 1 Recharge the battery before use 2 Clean the device by adding water to the Medicine Cup press the power button for 3 seconds then...

Страница 14: ...red equipment Nebulization Rate 0 2ml min minimum Particle Size 2 2 m 25 Medicine Capacity 8ml maximum 0 5ml minimum Battery Life the battery can be used for approximately 8 days after charging if ope...

Страница 15: ...Warning Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary this equipment and the other equipme...

Страница 16: ...issions are very low and are not likely to cause any interference in nearby electronic equipment The Mesh Nebulizer with Rechargeable Battery is suitable for use in all establishments including domest...

Страница 17: ...ectromagnetic environment guidance 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV for power supply lines 1 kV differential mode 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 7...

Страница 18: ...fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked wit...

Страница 19: ...hat to which the receiver is connected Consult the dealer or an experienced for help This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This devi...

Страница 20: ...Name Tel Sales shop name Sales shop address Address Zip code Product model Date of purchase Dealer info Mesh Nebulizer with Rechargeable Battery Service Warranty Card...

Страница 21: ...WARNING No modifications of this equipment are allowed...