14
15
PUSH
Picture 8 remove accessories
Picture 9 disassemble the Medicine Cup
Picture 10 after using
6. After Using
(2) To remove the Medicine Cup from the main device,press the button at the rear of
the Device and push Medicine Cup forward.
(1) Remove the mouthpiece or mask
(3) Open the lid, pour out all the residual liquid, and promptly clean and disinfect the
device, allow all the parts to air dry and store the device in a clean environment
.
See Chapter 5 for more information.
5. Turn Off Power
Press the Power button to turn off the power. The Power Indicator (Green Light) will
go out.
1. Cleaning
Chapter 5 Maintenance of the Nebulizer
Be sure to turn off the device.
After each use, first discard all residual medicine, and remove the attachment
from the Main Unit to clean and disinfect. Dry the parts immediately after washing.
Do not store when Main Unit and Medicine Cup have any remaining liquid or moisture.
1.
Medicine Cup
1.1 Take the Medicine Cup off the nebulizer, open the cover and pour out the residual liquid.
1.2 Add enough purified water (3-6ml) to the Medicine Cup and then close the cover tightly.
Shake the cup gently to dissolve any residual medication with purified water.
1.3 Open the cover of Medicine Cup and pour our all the liquid.