150kHz to 80MHz

d=[3.5/V1]

0.01

0.035

0.07

/

0.1

0.11

0.22

/

1

0.35

0.7

/

10

1.11

2.21

/

100

3.5

7

/

80MHz to 800MHz

80MHz to 2.7GHz

Recommended separation distances between portable and mobile RF 

communications equipment and the EQUIPMENT or SYSTEM -for EQUIPMENT 

and SYSTEMS that are not LIFE-SUPPORTING

The Mesh Nebulizer with Rechargeable Battery is intended for use in an 
electromagnetic environment in which radiated RF disturbances are controlled. 
The customer or the user of the Mesh Nebulizer with Rechargeable Battery can 
help prevent electromagnetic interference by maintaining a minimum distance 
between portable and mobile RF communications equipment (transmitters) and 
the Mesh Nebulizer with Rechargeable Battery as recommended below, according 
to the maximum output power of the communications equipment.

For transmitters rated at a maximum output power not listed above the 
recommended separation distance d in metres (m) can be estimated using the 
equation applicable to the frequency of the transmitter, where P is the maximum 
output power rating of the transmitter in watts (W) according to the transmitter 
manufacture

Separation distance according to frequency of 

transmitter(m)

Rated maximum 

output of 

transmitter(W)

where p is the maximum 

output power rating of the 

transmitter in watts (W) 

according to the transmitter 

manufacturer and d is the 

recommended separation 

distance in metres (m).b
Field strengths from fixed RF 

transmitters, as determined 

by an electromagnetic site  

survey,a should be less than 

the compliance level in each 

frequency range.
Interference may occur in the 

vicinity of equipment marked 

with the following  symbol:

NOTE1:At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE2:These guidelines may not apply in all situations. Electromagnetic is 

affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cord-

less) telephones and land mobile radios, amateur radio, AM and FM radio broadcast 

and TV broadcast cannot be predicted theoretically with accuracy. To assess the 

electromagnetic environment due to fixed RF transmitters, an electromagnetic site 

survey should be considered. If the measured field strength in the location in which 

the Mesh Nebulizer with Rechargeable Battery is used exceeds the applicable RF 

compliance level above, The Mesh Nebulizer with Rechargeable Battery should be 

observed to verify normal operation. If abnormal performance is observed, additional 

measures may be necessary, such as reorienting or relocating the Mesh Nebulizer with 

Rechargeable Battery.

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 

3V/m.

Radiated RF 

IEC 61000-4-3

EN 60601 

test level

Compliance 

level

Electromagnetic 

environment - guidance

10V/m

80MHz to 2.7GHz

10V/m

80MHz to 2.7GHz

Immunity test

30

31

Recommended separation distances between

portable and mobile RF communications equipment and 

the Mesh Nebulizer with Rechargeable Battery

d=[3.5/E1]

d=[3.5/E1]

d=[7/V1]

d=[7/V1]

Содержание NEB401

Страница 1: ...uction Manual Mesh Nebulizer with Rechargeable Battery NEB401 Rx Only Manufactured for Sunset Healthcare Solutions 180 N Michigan Ave Ste 2000 Chicago IL 60601 www sunsethcs com NEB401 IFU V1 01 2020...

Страница 2: ...nt the device may be used with pediatric 4 years of age and adult patients in the home hospital and sub acute care settings It is not intended for use with Pentamidine Table of Contents Chapter 1 Befo...

Страница 3: ...ooting Chapter 7 Customer Service Electromagnetic Compatibility Notes on FCC certification Chapter 9 Electromagnetic Compatibility Symbol Glossary Chapter 10 Symbol Glossary Cleaning Disinfecting Clea...

Страница 4: ...with pediatric 4 years of age and adult patients in the home hospital and sub acute care settings It is not intended for use with Pentamidine 1 What is the Mesh Nebulizer with Rechargeable Battery 2...

Страница 5: ...nvironment Warning For type or dose of medicine please follow the instructions of your physician and or licensed healthcare practitioner Important Safety Information Read Before Use The Mesh Nebulizer...

Страница 6: ...on in the treatment area such as sores herpes simplex or open wounds If you are not sure the device is safe for you or not please consult your doctor PRECAUTIONS Take the following precautions to avoi...

Страница 7: ...e end of USB cord to an adapter or other USB port such as USB port on a computer This nebulizer is powered with a built in lithium battery The battery can be charged with the USB cord 1 Power Supply C...

Страница 8: ...ed in the right place the device may not be able to operate normally NOTE Picture 5 Mask Assembly and Mouthpiece Assembly 4 Usage of Nebulizer Nebulizer Accessories are categorized as mouthpiece adult...

Страница 9: ...hile nebulizing 1 If this nebulizer is being used to treat a life threatening condition a backup device is necessary 2 The nebulizer may be unable to be used due to power failure or mechanical malfunc...

Страница 10: ...n off the power The Power Indicator Green Light will go out 1 Cleaning Chapter 5 Maintenance of the Nebulizer Be sure to turn off the device After each use first discard all residual medicine and remo...

Страница 11: ...r 3 6 ml 1 5 Press and hold the Power button for 3 seconds WHILE THE DEVICE IS OFF this will turn on the light and alternate between green and blue colors Release the button allowing the device to ent...

Страница 12: ...utes for each time 3 years for main unit 3 months for Medicine Cup For the battery the device can be used for approximately 8 days after charging if operating for 30 minutes a day Lithium battery has...

Страница 13: ...se under guidance of doctor s prescription The device is on but weak 1 Recharge the battery before use 2 Clean the device by adding water to the Medicine Cup press the power button for 3 seconds then...

Страница 14: ...red equipment Nebulization Rate 0 2ml min minimum Particle Size 2 2 m 25 Medicine Capacity 8ml maximum 0 5ml minimum Battery Life the battery can be used for approximately 8 days after charging if ope...

Страница 15: ...Warning Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary this equipment and the other equipme...

Страница 16: ...issions are very low and are not likely to cause any interference in nearby electronic equipment The Mesh Nebulizer with Rechargeable Battery is suitable for use in all establishments including domest...

Страница 17: ...ectromagnetic environment guidance 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV for power supply lines 1 kV differential mode 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 7...

Страница 18: ...fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked wit...

Страница 19: ...hat to which the receiver is connected Consult the dealer or an experienced for help This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This devi...

Страница 20: ...Name Tel Sales shop name Sales shop address Address Zip code Product model Date of purchase Dealer info Mesh Nebulizer with Rechargeable Battery Service Warranty Card...

Страница 21: ...WARNING No modifications of this equipment are allowed...

Страница 22: ......

Отзывы: