4-12
Touch
“
MENU
”
( ) to go to
Menu screen.
Maintenance settings
Page
Maintenance
1/2
<Conditioning Arc>
After the electrodes are replaced, this function is used to condition a new electrode.
The number of conditioning arc is preset. Arc Count is automatically reset after
conditioning arc is performed.
<Reset Arc Count>
Arc Count can be reset, while All Count cannot.
2/2
<Restore Data>
All parameters except Arc Count and All Count are returned to the factory setting.
<Self Inspection>
The circuit board, optical unit, motors and heat shrink ovens are automatically
inspected.
<Touch Calibration>
Perform touch calibration when icons and items could not be selected correctly with
touching operation.
1/2
Maintenance
Reset Arc Count
Bright
Conditioning Arc
Arc Count
All Count
0100
00100
Turn power off and
Take power unit out of splicer
When electrodes changed
MENU
In the menu screen, touch
“Maintenance” icon ( ) to go to
Maintenance screen.
Touch the item once. The icon will
change its color. Then touch it again
to determine selection.
Touch the “move” ( , ) to go
next page.
2/2
Maintenance
Restore Data
Self Inspection
MENU
SM G652 Std.
60mm 0.25
Start
Ready
X/Y
DISP
Menu
Splice
Data
Maintenance
Security
Function
USB
Login
2011/01/01 00
:
00
DATA
Select Item
Information
Touch Calibration