TFG2900A Series Function/Arbitrary Waveform Generator Users’ Guide
Suin Instruments Co., Ltd
74
When using the amplitude coupling function, the instrument will set the amplitude or offset to the
limit value if the amplitude or offset of the target channel exceeds the upper limit or lower limit,
and the above formula is no longer applied to.
3.18.7 Coupling parameter
Press
〖
MultipleParam
〗
soft key to enter the coupling parameter setting menu. Four parameters
can be set: frequency ratio, frequency difference, amplitude difference and amplitude ratio.
3.18.8 Phase alignment
Press
〖
AlignPhase
〗
soft key and select the channel to be alignment, and press
〖
ConfirmAlign
〗
soft key, the synchronization signal of the selected channel can be aligned with the CH1 phase. At
this moment, the phase difference between each channel could be calculated according to phase
setting for each channel.
3.19 File Manager
File managers are used in the operations of harmonic synthesis, frequency sequences, arbitrary
waveforms, arbitrary wave sequences, instrument state, system update and USB disk operation.
File manager allows user to handle file save and recall conveniently in the above function.
3.19.1 Layout
The layout of the file manager is as below picture shown:
1. Current directory area: to display current opened directory (folder), that is the upper directory of
sub-directory display area.