TFG2900A Series Function/Arbitrary Waveform Generator Users’ Guide
Suin Instruments Co., Ltd
58
Move the X coordinate of the cursor to 26, press
〖
Point Remove
〗
soft key 10 times conti
nuously, the 10 points from 26 to 36 will be deleted as below picture shown.
3.16.5 Line Edition
Use ‘point edit’ to set voltage for each point in one waveform. But it will be inconvenien
t when there are lots of wave points need to be set. Then line edition can solve the prob
lem. Under this function, user only needs to set two points, then all points between the t
wo points will be automatically set by instrument according to linear relationship. It is mo
re convenient than point edition function and is applied to make linear modification for ex
isting waveform, or create a waveform consisting of broken lines.
For example, make a linear modification for a Sine with 100 points.
Firstly insert a Sine with 100 points and amplitude 1Vpp.