TFG2900A Series Function/Arbitrary Waveform Generator Users’ Guide
Suin Instruments Co., Ltd
65
Press
〖
Point
〗
soft key to set the horizontal coordinates of the mark point to 402. Then i
nstrument will zoom in taking horizontal axis as direction and mark point as center. Press
〖
X Zoom
〗
soft key and set the zoom ratio to 200%, waveform point (402,4.000V) wil
l be displayed as below picture shown.
Continue to increase the zoom ratio to 500%, and waveform point (400,4.000V) will displ
ay more clearly and reset the zoom ration to 200%.
Vertical Zoom: Press
〖
Point
〗
soft key and set the horizontal coordinate of mark point
to 260. Then observe another waveform point (251,550mV) just modified. The Sine is co
mpressed vertically because of larger amplitude of waveform point (400,4.000V), which le
ads to the waveform point (251,550mV) is barely visible, so vertical zoom function is req
uired in this situation.
Press
〖
Y Zoom
〗
soft key and select ‘vertical’, the instrument will zoom in taking
vertical axis as direction and mark point as center. Set the zoom ratio to 500% as below
picture shown.