SA5052 Digital Multimeter User ‟s Guide
Shijiazhuang Suin Instruments Co., Ltd. 52
DEF}
Measure the AC voltage as specified range and resolution and send the readings to outpu
buffer. For this measurement, the resolution is fixed as 5 1/2 digits, and this paramenter only
effects the front panel displaying.
MEASure : CURRent:DC?{<range>|MIN|MAX|DEF}
,
{<resolution>|MIN| MAX|
DEF }
Measure the DC voltage as specified range and resolution and send the readings to outpu
buffer.
MEASure:CURRent:AC?{ <range >|MIN|MAX|DEF}, (<resolution>|MIN|MAX|DEF}
Measure the AC current as specified range and resolution and send the readings to outpu
buffer. For this measurement, the resolution is fixed as 5 1/2 digits, and this paramenter only
effects the front panel displaying.
MEASure:RESistance? { <range>|MIN|MAX|DEF}
,
(<resolution >|MIN | MAX| DEF }
Measure the 2-wire resistance as specified range and resolution and send the readings to
outpu buffer.
MEASure:FRESistance? { <range >|MIN|MAX|DEF}
,
(<resolution>|MIN |MAX |DEF}
Measure the 4-wire resistance as specified range and resolution and send the readings to
outpu buffer.
MEASure:FREQuency?{<range>}MIN|MAX|DEF}
,
(<resolution>|MIN|MAX|DEF }
Measure the frequency as specified range and resolution and send the readings to outpu
buffer. For this measurement, multimeter select one range for all input signal between 3Hz to
900KHz. When there is no input, the measurement result is 0.
MEASure:PERiod?{
《
range >|MIN|MAX|DEF }
,
{<resolution>|MIN|MAX|DEF}
Measure the period as specified range and resolution and send the readings to outpu buffer.
For this measurement, multimeter select one range for all input signal between 1us and 0.33s.