SA5052 Digital Multimeter User ‟s Guide
Shijiazhuang Suin Instruments Co., Ltd. 59
by remote interface, the multimter will clear this value, for the value is stored in volatile
memory.
CALCulate:NULL:OFFSet{<Value>|MIN|MAX}
Save the zero value in zero register. Before writing in calculation register, you must enable
the calculation first and the zero setting can be set within the range from 0 to
士
120 of full
range. MIN=Full range
一
120% a MAX= 120% of full range, the setting is stored in volatile
memory.
CALCulate:NULL:OFFSet? [MIN|MAX}
Query zero value.
CALCulate:DB:REFerence{<Value>| MIN|MAX}
Save the dB value in register. Before writing in calculation register, you must enable the
calculation first and the dB setting can be set within the range from 0dBm to
士
200dBm.
MIN= - 200.00dBm MAX=200.00dBm, the setting is stored in volatile memory.
CALCulate:DB:REFerence?[MIN|MAX]
Query dB value.
CALCulate:DBM:REFerence{ <Value>|MIN|MAX}
Select dBm reference from the range 50 to 8000, MIN=50Ω, MAX=8000Ω, the setting is
stored in volatile memory.
CALCulate:DBM:REFerence{ MIN | MAX}
Query the dBm reference resistance.
CALCulate:LIMit:LOWer{<Value>|MIN|MAX}
Set the higher/lower limit for limit calculation within the range from 0 to
士
120% of full
range, the setting is stored in volatile memory.
CALCulate:LIMit:LOWer{MIN|MAX}
Query the lower limit.