SA5052 Digital Multimeter User ‟s Guide
Shijiazhuang Suin Instruments Co., Ltd. 67
Input protection: 10A 250V fuse
2-4 wire Resistance
Range
Full Scale
(1)
Resolution
Pass Current
5 1/2
4 1/2
100Ω
1kΩ
10kΩ
100kΩ
1MΩ
10MΩ
100MΩ
119.999Ω
1.19999kΩ
11.9999kΩ
119.999kΩ
1.19999MΩ
11.9999kΩ
99.9999kΩ
1mΩ
10mΩ
100mΩ
1Ω
10Ω
100Ω
1kΩ
10mΩ
100mΩ
1Ω
10Ω
100Ω
1kΩ
1kΩ
1mA
1mA
100µA
10µA
5µA
0.5µA
0.5µA
(1) 20% overranged value could be read except of range of 100MΩ
Accuracy
Speed
Range
(1)
1year 23±5
℃
(3)
S
100Ω
1kΩ
10kΩ
100kΩ
1MΩ
10MΩ
100MΩ
0.03+10
(2)
0.02+6
(2)
0.02+6
(2)
0.02+6
0.02+6
0.25+6
1+10
M
100Ω
1kΩ
10kΩ
100kΩ
1MΩ
10MΩ
100MΩ
0.03+16
(2)
0.02+8
(2)
0.02+8
(2)
0.02+8
0.02+8
0.25+16
1+20
F
100Ω
1kΩ
10kΩ
100kΩ
1MΩ
10MΩ
100MΩ
0.05+8
(2)
0.02+8
(2)
0.02+8
(2)
0.02+8
0.02+8
0.25+8
1+10
(1) To avoid the nosie interference to test lead, the shield leads are suggested for
resistance measurement above 100kΩ.
(2) relative zero mode (ZERO)