26.All elements of the selected parameter have been programmed. Turn the Key to the LOCK
position to leave the Programming Mode and re-enter the Operating Mode. The display will
then show the parameter and the number of fasteners tightened in the batch so far (0), in the
format "A. 0".
27.To set the elements for another parameter, rotate the Parameter Selection Switch to the
desired parameter, and return to Step 2 of this procedure.
NOTES:
It is not necessary to program every element to change one element of a parameter after
the first setup. Simply enter the PROG (programming), then use the UP keyswitch to move to
the element to be changed. Use the SET keyswitch to enter the programming of that element,
and the UP/DOWN/SET keyswitches as necessary to set the element to the desired value.
Press the SET keyswitch at the end to lock it in. If more elements are to be changed, use the
UP keyswitch to advance to the element in the menu, and program it. When the necessary ele-
ments have been changed, advance the elements until TBB is again displayed, then turn the
KEY to LOCK to resume operation.