Page 25
8. When the PTV receives a signal from the wrench, it will display first the channel on which it
received the signal, then the strength of the signal ("
GOOD
").
9. Press the UP keyswitch to get to the Time Between Batches. The LED Display will show:
TBB
10.Press the SET Key to enter the programming of the TBB. The LED Display will show "
0000
",
with the leftmost zero blinking. The four digits are the time in seconds that is the minimum
time required between batches of fasteners. To change the blinking number, use the DOWN
or UP arrow until the desired number is displayed and blinking. To accept the number that
you have selected, press the SET Keyswitch. This will lock in the number accepted, and
cause the next number to the right to become the active number.
11. Continue programming the TBB using the UP, DOWN, and SET keyswitches until the number
of seconds required between batches is displayed. Press the SET keyswitch again to lock in
the time. The unit will return to displaying "
TBB
".
12.Press the UP keyswitch to advance to the next element of the parameter. The LED Display
will then show "
TBC
", for Time Between Clicks. This is the minimum required time between
each wrench cycle (click) that the unit will require before accepting another signal.
13.Press the SET keyswitch to begin programming the TBC. Again the LED Display will display
"
0000
", with the leftmost zero the active one (the active digit is always the blinking one). The
time increment is seconds; the programming is to set the number of seconds required
between clicks. Use the UP, DOWN, and SET keyswitches to select then set each digit,
working from the 1000's column on the left to the single digit column on the right.
14.When the last digit of the TBC has been set, the LED panel will again display "
TBC
". Press
the UP keyswitch to advance to the next element of the parameter.
15.The next element is the TMAX or Time Maximum. This is the specification for the maximum
duration of a signal; the most time the wrench may be in the "clicked" position. To begin set-
ting this time, press the SET keyswitch.
16.The LED Display will show the current maximum time setting in a "seconds.tenths" format,
with a default setting of "
0.5
", or ½ second. The whole seconds digit (0) will be flashing as
the active digit. To change the value of the whole seconds part of the time, use the UP or
DOWN keyswitch. When the value desired is displayed, press the SET keyswitch to set the
value and move to the tenths part of the TMAX element. Use the UP and DOWN keyswitch-
es to display this portion of the element, then press the SET keyswitch to lock it in. Press the