Technical Data
43
EN
12
Technical Data
Power connection:
100-240 V~
Frequency:
50/60 Hz
Mains fuse:
2x T3, 15 A, UL-recognized
Max. power consumption:
82 VA
Max. current consumption:
100 V: 880 mA
240 V: 340 mA
Protection class:
I
Protection level:
BF
Moisture protection:
IP41 (Pump unit)
Moisture protection:
IP21 (Weight unit)
Classification:
IIb
Grounding:
Potential equalization plug
The device complies with the
following standards:
IEC 60601-1:1998 +A1:1991+A2:1995
IEC 60601-1:2005+CCInt1:2008+Int2:2009
EMC:
EN 60601-1-2 / IEC 60601-1-2
Operating conditions:
10-40 °C / 50-104 °F
30 to 75% rel. humidity
Storage and transportation
conditions:
-40 to +70 °C /-40 to +158 °F
10 to 90% rel. humidity
Dimensions:
Width x height x depth
710 x 1400 x 810 [mm]
28 x 55 x 32 [inch]
Mass:
26 kg /10.36 lb
Max. weight:
70 kg /10.36 lb
Load capacity:
0-40 kg
Max. load of weighing system:
50 kg
Maximum output pressure:
200 mm Hg
Adjustable pressure range:
15 to 100 and 100 to 150 mm Hg
Accuracy of pressure measurement:
± 1,5 measure value
Max. flow:
500 ml/min
Accuracy of flow measurement:
+10%/-3% measure value
±3% measure value
Accuracy of deficit display:
±10ml, ±3% of capacity, at 21 °C/ 69.8 °F
Connections/ports:
Data inputs/outputs
Содержание FLUID SAFE
Страница 1: ...EN ES FR DE Manual Manual Manuel Handbuch 2016 09 www stryker com 1000 400 948 Rev F ...
Страница 6: ......
Страница 8: ......
Страница 55: ...Appendix 49 EN 15 Appendix 15 1 Test Record Date Result Comment Signature ...
Страница 58: ......
Страница 60: ......
Страница 109: ...Apéndice 49 ES 15 Apéndice 15 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 114: ......
Страница 162: ...Annexe 48 FR 15 Annexe 15 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 166: ......
Страница 168: ......
Страница 216: ...Anhang 48 DE 15 Anhang 15 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Страница 220: ......
Страница 221: ......