Operation
Locking and unlocking the optional pendant control (Continued)
A
B
C
Button
Name
Function
A
Max Inflate
Inflates the support surface to
maximum pressure (60 mmHg)
B
Stop
Press once to return to previous
firmness setting
C
Lock Out / Unlock
Press to lock all pendant functions
Press again to unlock all pendant
functions.
Adjusting firmness with the optional pendant control
You can adjust the support surface firmness settings for patient comfort. The product has five levels of firmness
settings that range from 14 mmHg to 30 mmHg. The default of 22 mmHg provides optimal pressure relief for patients up
to 200 lb. For average sized patients, 18-22mmHg is recommended.
Notes
•
For larger patients, higher settings are recommended.
The increase firmness (+) and decrease firmness (-) buttons are inactive when Turn Assist or Max Inflate is active.