Operation
Locking and unlocking with the optional InTouch footboard (Continued)
A
B
C
Button
Name
Function
A
Max Inflate
Tap to inflate the support surface to
maximum pressure (60 mmHg).
B
Stop
Tap once to return to previous
firmness setting.
C
Lock Out / Unlock
Tap to lock all pendant functions.
Tap again to unlock all pendant
functions.
Adjusting firmness with the optional InTouch footboard
The support surface firmness settings may be adjusted for patient comfort requirements. The product has five levels of
firmness settings that range from 14 mmHg to 30 mmHg. The default of 22 mmHg will provide optimal pressure relief for
patients up to 200 lb. For average sized patients, 18-22mmHg is recommended.
Notes
•
For larger patients, higher settings are recommended.
•
Firmness increase and decrease are inactive when Turn Assist and Max Inflate are active.