Warning/Caution/Note Definition
Summary of safety precautions (Continued)
WARNING (CONTINUED)
•
Always inspect support surface covers (top and bottom) for tears, punctures, excessive wear, and misaligned
zippers each time the covers are cleaned. If compromised, the support surface cover should be removed from
service immediately and replaced to prevent cross-contamination.
•
Always perform preventative maintenance more frequently based on the usage level of the product. The life of the
support surface can be adversely affected by an increase in usage which may include more frequent cleaning and
disinfection.
•
Always allow the control box to completely dry before you place the support surface on top of it.
•
Always disinfect the support surface between patients, to avoid the risk of cross-contamination and infection.
•
Always make sure that you wipe each product with clean water and thoroughly dry each product after cleaning.
Some cleaning agents are corrosive in nature and may cause damage to the product if you use them improperly. If
you do not properly rinse and dry the product, a corrosive residue may be left on the surface of the product that
could cause premature degradation of critical components. Failure to follow these cleaning instructions may void
your warranty.
•
Always unplug the product power cord before cleaning or disinfecting to avoid the risk of shock.
CAUTION
•
Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in
this manual.
•
Do not modify the product or any components of the product. Modifying the product can cause unpredictable
operation resulting in injury to patient or operator. Modifying the product also voids its warranty.
•
Always lower the foot end section gently to avoid the risk of damage to the control box.
•
Do not allow sharp objects to come into contact with the support surface that could puncture tear or cut the cover.
•
Always secure all patient lines and tubing together before starting Turn Assist to prevent pulling, removal, or
breakage.
•
Do not iron, dry-clean, or tumble dry the support surface covers.
•
Do not power wash the support surface as this may cause malfunction and damage the product.
•
Always make sure that the support surface cover is completely dry before storing, adding linens or placing a patient
on the surface. Drying the product aids in preventing the performance of the product from being impaired.
•
Do not over expose the covers to higher concentration disinfectant solutions as these may degrade the covers.
•
Do not use accelerated hydrogen peroxides or quaternaries that contain glycol ethers as they may damage the
cover.