14
2920-109-001 REV G
www.stryker.com
Operation
Positioning the patient (Continued)
CAUTION
•
Do not allow sharp objects to come into contact with the support surface that could puncture tear or cut the cover.
To position the patient:
1. Place the patient in the center of the support surface, aligning the patient’s head toward the head board.
2. Check the patient frequently during Turn Assist and Lateral Rotation for proper positioning and support surface
inflation.
Activating CPR
When CPR is activated, the manifold bladder deflates, the pump is off and all of the valves are energized. CPR
activation is detected by a fast drop of pressure in a short period of time.
WARNING
Always make sure that the operator has access to the CPR straps.
Pull the red CPR strap (
) that is located on the left and right side of the head end of the mattress.
You can activate one or either of the straps.
Note:
The CPR visual alarm LED turns red. The CPR alarm LED continues to flash to indicate that a hazardous situation
has been detected. The CPR alarm LED will continue to flash until you reset the alarm.
Figure 13: CPR Strap
Resetting CPR with the optional pendant control
When you activate the CPR, the pendant CPR LED starts flashing red without an audible alert.
To reset the CPR:
1. Reinsert the red CPR straps by pulling down on the strap.
2. Press the
Stop
button and the Increase Firmness (
+
) button simultaneously to reset the CPR alarm. The support
surface will begin to reinflate.