![Stryker 2920 Скачать руководство пользователя страница 38](http://html.mh-extra.com/html/stryker/2920/2920_operation-manual_1385813038.webp)
34
2920-109-001 REV G
www.stryker.com
EMC information
(Continued)
Emissions test
Compliance
Electromagnetic environment
RF Emissions
CISPR 11
Group 1
PositionPRO
uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class A
PositionPRO
is suitable for use in all
establishments other than domestic and those
directly connected to the public low voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
Complies
WARNING
This equipment/system is intended for use by health care professionals only. This equipment/system may cause radio
interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such
as reorienting or relocating