Cleaning
WARNING
•
Always unplug the support surface power cord before cleaning or disinfecting to avoid the risk of shock.
•
Do not immerse support surface or foot box in cleaning or disinfectant solutions. To avoid the risk of product
damage or patient injury.
•
Do not allow fluid to pool on the support surface or foot box. Fluids can cause degradation of components and may
cause unpredictable safety and performance of this product.
•
Always inspect support surface covers (top and bottom) for tears, punctures, excessive wear, and misaligned
zippers each time the covers are cleaned. If compromised, the support surface cover should be removed from
service immediately and replaced to prevent cross-contamination.
•
Always perform preventative maintenance more frequently based on the usage level of the product. The life of the
support surface can be adversely affected by an increase in usage which may include more frequent cleaning and
disinfection.
CAUTION
•
Do not iron, dry-clean, or tumble dry the support surface covers.
•
Do not power wash the support surface as this may cause malfunction and damage the product.
•
Always make sure that the support surface cover is completely dry before storing, adding linens or placing a patient
on the surface. Drying the product aids in preventing the performance of the product from being impaired.
Always follow hospital protocol for cleaning and disinfecting.
To clean the support surface covers between patient uses, follow these steps in order:
1.
Unplug the support surface before cleaning.
2.
Lift up the foot section of the support surface to clean the bottom surface.
3.
Using a clean, soft, damp cloth, wipe the support surface covers with a mild soap and water solution to remove
foreign material.
4.
Wipe the support surface covers with a clean, dry cloth to remove any excess liquid or cleaning agent.
Cleaning the control box
WARNING
Always allow the control box to completely dry before you place the support surface on top of it.
To clean the control box:
1.
Unplug the support surface before cleaning.
2.
Using a clean, soft, damp cloth, wipe the control box with a mild soap and water solution.
3.
Wipe the control box with a clean, dry cloth to remove any excess liquid or cleaning agents.
4.
Thoroughly rinse and dry the control box after cleaning.
5.
Disinfect the control box with a hospital grade disinfectant as necessary.
Cleaning the pendant
1.
Unplug the pendant before cleaning.
2.
Using a clean, soft, damp cloth, wipe down the pendant using a mild soap and water solution to remove foreign
materials.
3.
Wipe down the pendant with a clean, dry cloth to remove any excess liquid or cleaning agents.