Preventive maintenance
Fire barrier replacement (Continued)
1.
Raise the bed height to the full up position.
2.
Lower the fowler and gatch sections to the full down positions.
3.
Unplug the support surface from the power source.
4.
Unplug one of the CPR plugs to deflate the bladder.
5.
Unzip the top cover and fold it off to the side.
6.
Pull the fire barrier off starting from the head end pulling down to the foot end, working it from the left to the right
while pulling it down.
7.
Discard the fire barrier.
8.
Starting at the foot end, roll the new fire barrier up and slide the fire barrier over the crib assembly.
Note:
Align the fire barrier on the crib before sliding over the crib assembly.
9.
Carefully slide the fire barrier up the crib assembly, working from side to side, to make sure that the fire barrier is
tight on the crib assembly.
10. Align the crib assembly on top of the bottom part of the cover.
Note:
Spread the excess fire barrier material equally below the crib assembly at the head end.
11. Fold and align the top cover over the top of the crib assembly.
12. Zip the cover to close. Start at the head end patient right corner and stop at the foot end patient right corner.
13. Verify proper operation before returning the product to service.