• Immediate-use steam sterilization (IUSS, or “Flash”) is intended only
for emergency situations. In the event that immediate-use steam
sterilization is required, the following instructions should be used.
Immediate-use steam sterilization should only be
used in carefully selected clinical situations (e.g. an
instrument needed for a case falls on the floor and no
replacement instrument is available). IUSS should not
be used due to inadequate inventory.
“Flash” Pre-Vacuum
Wrapping
None
Temperature
132 °C (270 °F)
134 °C (274 °F)
Sterilization Time
4 minutes
3 minutes
Dry Time
None
5. Allow the camera head, cable, coupler, and scope to completely dry
before reassembly. Any moisture on the threads will cause the camera
and coupler windows to fog during use.
• Drying time depends on several variables including: altitude,
humidity, type of wrap, preconditioning, size of chamber, mass of
load, material of load, and placement in the chamber. Users must
verify that drying time set in their autoclave yields dry surgical
equipment.
Steris/Amsco V-PRO
1. Clean and prepare the camera head and cable as recommended
in this user guide. Ensure the soaking cap is installed. Refer to the
“Installing the Soaking Cap” section for more detail about installing
the cap.
2. If using a sterilization tray (optional), follow any additional
instructions provided with the tray. Use only trays that are approved
for sterilization with V-PRO.
3. Double wrap the camera head and cable (or tray) prior to sterilization
4. Sterilize the device using one of the following V-PRO sterilization
systems:
• V-PRO maX Sterilizer (Non-Lumen or Standard cycle)
• V-PRO 1 Plus Sterilizer (Non-Lumen or Standard cycle)
• V-PRO 1 Sterilizer (Standard cycle)