Operation
Note:
Before operating the Precision AC Camera, ensure all system
components have been set up according to the instructions in the “Setup and
Interconnection” section.
Operating the Camera with a Light Source
IMPORTANT SAFETY NOTICE:
Before operating this device,
please read this operating manual thoroughly and carefully.
When using this device with a light source, fire and/or severe
injury may result to the patient, user or inanimate objects if
the instructions in this manual are not followed.
All light sources can generate significant amounts of heat
(exceeding 41 °C/106 °F) at the scope tip, the scope light post,
the light cable tip, and/or near the light cable adapter. Higher
levels of brightness from the light source result in higher levels
of heat. Always adjust the brightness level of the camera and
the monitor before adjusting the brightness level of the light
source. Adjust the brightness level of the light source to the
minimum brightness necessary to adequately illuminate the
surgical site.
In addition, adjust the internal shutter of the camera higher
in order to run the light source at a lower intensity. Avoid
touching the scope tip or the light cable tip to the patient,
and never place them on top of the patient, as doing so may
result in burns to the patient or user. In addition, never place
the scope tip, the scope light post, the light cable adapter, or
the light cable tip on the surgical drapes or other flammable
material, as doing so may result in fire.
Always place the light source in standby mode whenever
the scope is removed from the light cable or the device
is unattended. The scope tip, scope light post, light cable
adapter, and light cable tip will take several minutes to cool
off after being placed in standby mode, and therefore may still
result in fire or burns to the patient, user, or inanimate objects.