Drying
• For automated drying, use the drying cycle provided with the washer/
disinfector.
• For manual drying, use a clean, lint-free cloth.
• Filtered pressurized air (≤40 psi) can be used to aid in drying any
crevices, mated surfaces, and hard-to-reach areas.
Disinfection (Optional)
Thermal Disinfection
The device must be sterilized after disinfection. Failure
to sterilize the device before reuse presents an acute
infection control risk to the patient.
Follow the steps below to disinfect the device using thermal disinfection.
1. Place the device into the washer disinfector.
2. Program the washer using the following parameters:
Thermal Disinfection
Recirculation Time:
1 Minute
Water Temperature
90°C