40
Periodic Maintenance Schedule
To ensure safe operation of the Precision AC Camera, you
should periodically perform the following procedure:
Every 12 months, check the earth leakage current to <500 µA (<300 µA
in USA), ground protective earth impedance to <0.1 ohms, and power
consumption less than or equal to rated power. Use a true RMS digital
multimeter and safety analyzer to perform this test.
Note:
Refer calibration and operating difficulties not detailed in this manual
to your Stryker Endoscopy sales representative.
Expected Service Life
Both models of the Precision AC Camera Control Unit have an expected
service life of 5 years.
The Precision AC Camera Head has an expected service life of 2 years, except
with repeated use of an automated washer the expected service life is 1 year.
Note:
If the camera head is cleaned with the automated cleaning method,
steam sterilization is recommended. Repeated automated cleaning when
sterilizing with Sterrad or V-Pro can degrade the product’s cosmetic
appearance.
Disposal and Recycling Information
This product contains electrical waste or electronic equipment.
It must not be disposed of as unsorted municipal waste and
must be collected separately in accordance with applicable
national or institutional related policies relating to obsolete
electronic equipment.
Dispose of the camera system according to local laws and hospital practices.