16
Connecting the DVI Fiber Outputs
Using adjustments or performing procedures differently than
specified below may result in hazardous radiation exposure.
The Precision AC Camera has the optional upgrade, the fiber output, for
model 0700010001. This upgrade contains four laser diodes to transmit a DVI
output over fiberoptic cables.
1. Connect four individual fibers (terminated in Lucent connectors) to the
red (R), green (G), blue (B) and clock (C) laser diodes on the console rear
panel.
2. Connect the four fibers to a compatible fiberoptic DVI receiver.
• The four fibers should be connected to the camera console in the
labeled order: RGBC
• The fibers should be connected to the monitor in one of two
configurations: CBGR (reverse order) or BGRC (R/B switched).
Note:
The Precision AC Camera Control Unit, model 0700010001, is a Class 1
laser product per IEC 60825-1 and 21 CFR 1040.
Installing the Soaking Cap
Before reprocessing the camera head, the soaking cap must be installed to
avoid damaging the cable connector.
Caution:
Failure to properly tighten the soaking cap will corrode the
connector pins and void the warranty.
• To install the soaking cap, screw the cap onto the threads of the cable
connector until it forms a tight seal.
• To remove the soaking cap, unscrew the cap and pull it away from the
cable connector.