31
Instructions for Reprocessing
Point of Use
• Disassemble the camera head from the scope and coupler. To
disconnect the scope, depress the endobody clamp on the coupler and
remove the scope from the coupler. To disconnect the coupler, grip the
rear adapter of the coupler and unscrew it counterclockwise from the
camera head.
• Wipe any excess soil from the device using a clean sterile cloth.
Containment and Transportation
• Reprocess the device as soon as reasonably practical following use.
• Transport the device in a tray to avoid damage. Follow the facility’s
internal procedures for the transportation of contaminated surgical
instruments and devices.
Cleaning
Manual Cleaning
1. Soak
• Ensure the soaking cap is installed. Refer to the “Installing the
Soaking Cap” section for more detail about installing the cap.
• Prepare a solution of enzymatic detergent and tap water at
35 °C. Refer to the manufacturer’s instructions for detergent
concentration and any other recommendations.
• Fully immerse the device and use an appropriately-sized syringe to
inject any mated surfaces and crevices with at least 50 mL of the
prepared detergent.
• Soak the device for a minimum of 15 minutes.
2. Brush
• Thoroughly brush the exterior of the device with a soft-bristled
brush for 90 seconds, focusing on any crevices and mated surfaces.
• Inject any mated surfaces and crevices a minimum of 5 times with
50 mL of the prepared detergent.