29
Cautions
• Always install the soaking cap prior to processing the camera. Failure to
properly tighten the soaking cap will corrode the connector pins and
void the warranty. Refer to the “Installing the Soaking Cap” section for
more detail about installing the cap.
• Inspect the camera cable for cuts and breaks before soaking in any
fluid. Return any damaged camera to Stryker for service.
• Never soak the camera in the same tray with sharp instruments.
• Do not use brushes or pads with metal or abrasive tips during manual
cleaning, as permanent scoring or damage could result.
• To minimize galvanic corrosion, avoid soaking dissimilar metals in close
proximity.
• Allow the camera head to cool before connecting it to the console.
Connecting the camera head while it is hot may result in system error.
Limitations on Reprocessing
• Do not cross-sterilize the device. Using multiple sterilization methods
may significantly reduce the performance of the device.
• Do not leave the device in solutions longer than necessary. This may
accelerate normal product aging.
• Proper processing has a minimal effect on this device. End of life is
normally determined by wear and damage due to use.
• If the camera head is cleaned with the automated cleaning method,
steam sterilization is recommended. Repeated automated cleaning
when sterilizing with Sterrad or V-Pro can degrade the product’s
cosmetic appearance.
• Repeated sterilization via Sterrad 100NX can degrade the product’s
cosmetic appearance.
• Damage caused by improper processing is not covered by the
warranty.