ASTOPAD
®
Instructions for Use
23
6.4 Control panel
Fig. 3 Control panel
No. Designation
Description
1
Actual temperature
A or B
Displays the actual temperature of the applied part
2
Set temperature
A or B
Displays the selected set temperature of the applied
part.
3
“Temperature
increase” button
A or B
Press this button to increase the set temperature in
increments of 0.5 °C.
4
“Temperature
decrease” button
A or B
Press this button to decrease the set temperature in
increments of 0.5 °C.
5
Alarm indicator
Displays the corresponding alarm code in an alarm
situation.
6
“Start” button
“Start” LED (green)
A or B
Press this button to start the heating process
or
Press this button to confirm a change made to the set
temperature.
7
“Stop” button A or B
Ends the heating process and switches off the respec-
tive output.
8
Battery status
indicator
Indicates the current battery status.
9
“Standby” button
“Standby” LED (blue)
The Standby button switches between the
Standby
mode
and the
On mode
.
10
“Alarm” LED (yellow)
LED flashes or turns on and the acoustic alarm signal
sounds when there is an alarm situation.
11
Display
Informs the user of temperatures and fault conditions.
Содержание ASTOPAD
Страница 59: ...ASTOPAD Instructions for Use 59...