10.3 Time alarm “A3”
(low priority alarm condition)
Control panel
Device response
•
When the applied part is started, the display
shows
“A3”.
•
The Start LED
flashes and the Alarm LED
turns
on.
•
An acoustic alarm signal is triggered.
Alarm condition
This alarm is triggered if the set temperature
is not
reached during 60 minutes of uninterrupted heating.
Possible reasons
►Required
action(s)
A layer of thermally conductive materials (water or gel) is
on the applied part.
►Remove the layer or place it underneath the applied part.
Applied part defective.
►Send the applied part to the local sales point.
Required action(s)
for resetting
Start the heating process again with the “START”
but-
ton or if necessary, switch off the control unit with the
“Standby”
button.