ASTOPAD
®
Instructions for Use
31
WARNING
Risk of injury!
•
Do not use the ASTOPAD until the following error conditions have been rem-
edied through appropriate corrective action:
- Damaged or worn cables, plugs, or connecting socket.
- Damaged housing, damaged or loose control panel
- Damaged or missing markings/safety signs/warnings.
- Damaged outer cover of the applied parts.
- Alarm testing device defective. No visual or acoustic alarm after switching
on using the Standby button. (Self-test)
- Button(s) which do not function correctly.
- A system that has been exposed to mechanical impact or liquid ingress into
internal electronic system components.
- A system that has previously delivered an electric shock to a person.
- A system that appears to be overheating.
- If at least one of the applied parts in a system triggers or has triggered an
alarm shut-off.
WARNING
Risk of decubitus ulcer!
•
Regardless of the treatment duration, aged, paralyzed, comatose, and
cachectic patients are particularly at risk of decubitus ulcers. Critical
points should therefore also be constantly examined by medical personnel.
•
Do not fold, bend, or operate the applied parts in a folded condition.
•
Do not place the patient on the connection block of the applied part.
•
When the ASTOPAD COV applied parts are used as an under-blanket, en-
sure that they are placed flatly underneath the patient, are secured, and will
not crease.
•
Care must be taken during any surgical interventions to ensure that sufficient
measures are taken to prevent decubitus ulcers in accordance with the pa-
tient's positioning.
•
The risk of skin irritation caused by pooling of surgical prep solutions under
the patient may increase with warming; ensure that surgical prep solution in-
structions for use are followed.
•
Alleviate or remove the risk of heating skin under pressurized bony promi-
nences.
•
DO NOT place any hard objects (e.g., mattress cables, ECG cables, hard
cautery return pads, patient fluid lines, etc.) between the ASTOPAD applied
part and the patient.
•
ASTOPAD COV applied parts can be wrapped around the patient. However,
take care that the applied part will not form creases.