5-17
764333-675
5.4.6
Printer Prints:
TC3 FAILURE
Occurs if the drain block thermocouple
reading is greater than 75°C or less than
0°C in DIAGNOSTIC Cycle.
1. Check TC3 for loose
connections or broken
wires.
2. Check analog cable
connections. (Refer to
Figure 5-5.)
3. Re-enter known good
calibration values.
4. Defective TC3. Replace.
5. Defective interface board.
Replace board.
5.4.7
Printer Prints:
CHM-PRES
NOT
CALIBRATED
Indicates PT1 has not been calibrated or
calibration data lost and default values
are being used.
1. Calibrate the PT1.
2. Re-enter known good
calibration values.
3. Defective battery-backed
RAM. Replace RAM.
5.4.8
Printer Prints:
INLET NOT
CALIBRATED
Indicates TC1 (inlet thermocouple) has
not been calibrated or calibration data
lost and default values are being used.
1. Re-enter known good
calibration values
2. Calibrate TC1.
3. Defective battery-backed
RAM. Replace RAM.
5.4.9
Printer Prints:
OUTLET NOT
CALIBRATED
Indicates TC2 (outlet thermocouple) has
not been calibrated or calibration data
lost and default values are being used.
1. Re-enter known good
calibration values.
2. Calibrate TC2.
3. Defective battery-backed
RAM. Replace RAM.
5.4.10 Printer Prints:
DRN-BLK NOT
CALIBRATED
Indicates TC3 (drain block
thermocouple) has not been calibrated
or calibration data lost and default
values are being used.
1. Re-enter known good
calibration values.
2. Calibrate TC3.
3. Defective battery-backed
RAM. Replace RAM.
5.4.11 Printer Prints:
CONC. NOT
CALIBRATED
Indicates concentration monitor has not
been calibrated or calibration data lost
and default values are being used.
1. Calibrate the
concentration monitor.
2. Re-enter known good
calibration values.
3. Defective battery-backed
RAM. Replace RAM.
5.4.12 Printer Prints:
FLASH VALUES
RESTORED
Flash Memory Parameters Restored.
Occurs on power-up if battery-backed
RAM values are lost. RAM is initialized.
Flash ROM values are restored.
1. Battery Backed RAM
failure. Replace RAM.
2. Main control board failure.
Replace board.
5.4 CONTROL FAULTS (CONTINUED)
Fault
Description
Service Instructions
Содержание SYSTEM 1E
Страница 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Страница 3: ......
Страница 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Страница 72: ...5 20 764333 675 Figure 5 3 DC Input Harness Pin Out Locations Figure 5 4 DC Power Harness Pin Out Locations ...
Страница 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Страница 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Страница 108: ......
Страница 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Страница 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Страница 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Страница 123: ...14 764333 676 Figure 3 Frame Components Panels 2 1 13 22 21 10 11 10 9 8 6 24 23 5 4 3 7 24 12 ...
Страница 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Страница 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Страница 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Страница 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Страница 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Страница 166: ......
Страница 167: ......
Страница 168: ......
Страница 169: ......
Страница 170: ......