![Steris SYSTEM 1E Скачать руководство пользователя страница 97](http://html.mh-extra.com/html/steris/system-1e/system-1e_maintenance-manual_1374323097.webp)
7-20
764333-675
7.23 UV MONITOR REPLACEMENT
DANGER – HIGH VOLTAGE: Before
attempting to dismantle, isolate power at
control panel.
DANGER – ULTRA VIOLET LIGHT: UV
Light is harmful to eyes and exposed skin.
Do not run system with any covers or
fittings removed.
WARNING–PERSONAL INJURY and /or
EQUIPMENT DAMAGE HAZARD:
•
The UV System is not user serviceable. All
installation, maintenance work and service
must be carried out by an authorized STERIS
Service Representative only. Use only
STERIS-approved replacement lamps,
quartz sleeves and monitors.
•
UV chamber may be under pressure. Before
attempting to remove end plates, monitor or
quartz sleeve, depressurize and drain
chamber.
WARNING – POTENTIAL BURN HAZARD:
Possible high temperature. Before
attempting to dismantle, make sure UV
chamber is in a cool state.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE HAZARD: Only use STERIS
Corporation approved UV lamps and
Quartz sleeves.
UV Monitor:
1. Unplug UV System from facility GFCI receptacle.
2. Turn off incoming water supply to ultra violet
system.
3. Disconnect UV monitor cable from back of
processor. Allow system to cool for 10 minutes.
4. Loosen and unthread the brass nut from sensor
Quartz Hi-Hat assembly and remove sensor
assembly as shown in Figures 7-18 and 7-19.
5. Install both new O-rings onto the replacements
sensor assembly as shown in Figure 7-19.
Figure 7-18. Loosen Brass Nut
Brass
Nut
!
!
!
Figure 7-19. Install O-Rings
!
!
O-Rings
Содержание SYSTEM 1E
Страница 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Страница 3: ......
Страница 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Страница 72: ...5 20 764333 675 Figure 5 3 DC Input Harness Pin Out Locations Figure 5 4 DC Power Harness Pin Out Locations ...
Страница 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Страница 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Страница 108: ......
Страница 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Страница 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Страница 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Страница 123: ...14 764333 676 Figure 3 Frame Components Panels 2 1 13 22 21 10 11 10 9 8 6 24 23 5 4 3 7 24 12 ...
Страница 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Страница 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Страница 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Страница 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Страница 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Страница 166: ......
Страница 167: ......
Страница 168: ......
Страница 169: ......
Страница 170: ......