5-15
764333-675
5.3 OTHER FAULTS
5.3.1
Ground Fault Interrupt (GFI) Trip
Description
– A GFI measures the difference in
current flow between the hot and neutral lines of a
circuit. Normally both are identical; however, if there
is current leakage to ground from the hot line, the GFI
senses the imbalance and
trips
.
The typical trip point of a GFI is 4 to 6 mA, and since
this is a very small level of imbalance, it will require
component isolation to locate the cause.
Establish if the GFI trips as soon as the unit turns on,
or if it occurs in cycle. If trip occurs when turned on, it
may be necessary to isolate components one at a time.
When doing so, turn the unit ON to see if the GFI trips.
If trip occurs in cycle, the trip may be occurring when
a specific component is energized; if so, use the service
mode/hand terminal to manually activate individual
components until the source of trip is located.
Action:
1. Verify proper operation of GFI using ground fault
tester – if OK, continue with component isolation,
otherwise, replace GFI.
2. If trip occurs in cycle – use service mode/hand
terminal to operate components individually until
trip source is located.
3. If trip occurs when powered up – isolate individual
components manually until fault is located.
5.3.2
Water Spills from SYSTEM 1E Processor
Action:
Check the following items.
1. Drip pan full of water?
2. Is chamber filling?
a.
Faulty LS3 – repair.
b.
SOL5 or SOL7 leaking – repair.
c.
Are instruments loaded incorrectly – reload.
d.
Is water pressure >60 psi – reduce.
e.
Back pressure in the drain – repair.
f.
Faulty CK2, CK3 – repair.
g.
Header block – replace lid.
Содержание SYSTEM 1E
Страница 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Страница 3: ......
Страница 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Страница 72: ...5 20 764333 675 Figure 5 3 DC Input Harness Pin Out Locations Figure 5 4 DC Power Harness Pin Out Locations ...
Страница 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Страница 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Страница 108: ......
Страница 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Страница 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Страница 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Страница 123: ...14 764333 676 Figure 3 Frame Components Panels 2 1 13 22 21 10 11 10 9 8 6 24 23 5 4 3 7 24 12 ...
Страница 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Страница 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Страница 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Страница 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Страница 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Страница 166: ......
Страница 167: ......
Страница 168: ......
Страница 169: ......
Страница 170: ......