7-13
764333-675
4. Slip used ink cartridge off end of paper and discard.
5. Slip new ink cartridge over paper, in same position
as one removed, making sure paper feeds through
printer and through space between ink cartridge
ribbon and cartridge housing.
6. Press both left and right sides of cartridge
simultaneously until cartridge snaps into place.
7. Tighten any slack in ink ribbon by turning
adjustment wheel counterclockwise approximately
1/4-turn.
8. Close housing of printer take up.
7.19 CHANGING MAXPURE™ FILTER
WARNNG – PERSONAL INJURY AND/OR
EQUIPMENT DAMAGE HAZARD: SYSTEM
1E Processor performance has been
validated and optimized with components defined
by STERIS in the Operator Manual for the SYSTEM
1E Processor. Use of pre-filters that do not meet
STERIS's specifications can lead to premature
failure of the MaxPure Filter inside the processor.
STERIS’s warranty will not apply to damage or early
life failure of the MaxPure Filter resulting from the
use of pre-filters that do not meet STERIS’s
specifications. DO NOT USE components that do
not meet STERIS’s specifications.
Procedure:
MaxPure Filter
- The MaxPure Filter is replaced, at a
minimum, once every 90 days. The filter is also
changed if the processor indicates a MaxPure Filter
Membrane Test failure (see
S
ECTION
5,
T
ROUBLESHOOTING
) or a Fill Time problem.
If the MaxPure Filter has not been changed within
90 days, the processor will printout the following
message:
WARNING
MAXPURE FILTER
IN USE 90 DAYS
SEE OPERATOR MANUAL
A Liquid Chemical Sterilant Processing Cycle or a
Diagnostic Cycle cannot be started until the MaxPure
filter is changed. If a cycle is attempted, the display for
the System 1E processor will show:
Change MaxPure
Filter
1. Unplug SYSTEM 1E Processor.
2. Locate housing of MaxPure Filter.
Open control panel.
Locate housing of MaxPure Filter.
!
Содержание SYSTEM 1E
Страница 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Страница 3: ......
Страница 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Страница 72: ...5 20 764333 675 Figure 5 3 DC Input Harness Pin Out Locations Figure 5 4 DC Power Harness Pin Out Locations ...
Страница 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Страница 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Страница 108: ......
Страница 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Страница 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Страница 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Страница 123: ...14 764333 676 Figure 3 Frame Components Panels 2 1 13 22 21 10 11 10 9 8 6 24 23 5 4 3 7 24 12 ...
Страница 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Страница 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Страница 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Страница 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Страница 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Страница 166: ......
Страница 167: ......
Страница 168: ......
Страница 169: ......
Страница 170: ......