![Steris SYSTEM 1E Скачать руководство пользователя страница 31](http://html.mh-extra.com/html/steris/system-1e/system-1e_maintenance-manual_1374323031.webp)
2-11
764333-675
• Pressure Calibration
Select
PRESS
, then press
SAVE VALUES
touch pad.
Display shows:
WARNING! CALIB
WILL BE CHANGED
Press
SAVE VALUES
to continue. Display shows:
INFLATING SEAL
Display shows:
FILLING CHAMBER
Display shows:
STABILIZING P=
0.XX psi
The control holds for a period of time, allowing
pressure to stabilize, read the pressure on the
transducer due to the head of water in the chamber,
and set the calibration point automatically. No
adjustments are necessary by the technician. Unit
proceeds to drain. Display shows:
DRAINING CHAMBER
Display shows:
* UPDATING *
* MEMORY *
Display shows:
DEFLATING SEAL
Display shows:
* PRESSURE *
* CALIBRATED *
Printer prints pressure-calibrated values (see
Figure 2-6, for example). Press
SAVE VALUES
touch
pad to continue.
• Temperature Calibration
Temperature calibration is not normally necessary on
SYSTEM 1E Liquid Chemical Sterilant Processing
System.
NOTE: A calibrated temperature measuring device is
required for this procedure.
A container with cold water (5 to 10°C [41 to 50°F]) and
hot water (55 to 65°C [131 to 149°F]) is necessary. The
thermocouple(s) is removed from the mounting
location and is placed in the bath.
NOTE: To make calibration easier, use a spare
thermocouple in place of TC1, TC2 and TC3
thermocouples in the unit. Calibrate each
thermocouple separately. Attach the spare
thermocouple to the proper connection on CN1 while
leaving the threaded end hang free. Free-hanging end
is inserted into cold and hot water baths as directed
during procedure.
=====================
======PRESSURE=======
=====CALIBRATED======
=====================
AT X:XX:XX ON XX-XX-XX
PREVIOUS VALUES
PRESSURE TRANSDUCER
LOW RAW =XXXXX
CALIBRATED VALUES
PRESSURE TRANSDUCER
LOW RAW =XXXXX
=====================
Figure 2-6. Pressure
Calibrated Values
Содержание SYSTEM 1E
Страница 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Страница 3: ......
Страница 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Страница 72: ...5 20 764333 675 Figure 5 3 DC Input Harness Pin Out Locations Figure 5 4 DC Power Harness Pin Out Locations ...
Страница 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Страница 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Страница 108: ......
Страница 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Страница 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Страница 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Страница 123: ...14 764333 676 Figure 3 Frame Components Panels 2 1 13 22 21 10 11 10 9 8 6 24 23 5 4 3 7 24 12 ...
Страница 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Страница 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Страница 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Страница 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Страница 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Страница 166: ......
Страница 167: ......
Страница 168: ......
Страница 169: ......
Страница 170: ......