Signal Input
Input [Select]
The Input [Select] key selects between the voltage and current preamps.
The voltage inputs (A and B) have a 10 MΩ, 25 pF input impedance.
Their connector
shields are isolated from the chassis by either 10
Ω (Ground) or 10
kΩ (Float). Do not
apply more than 10 V to either input. The shields should never exceed 1 V.
The current input uses the I connector. The input burden resistance is 100
Ω (
1 µA range)
or 1 k
Ω (1
0 nA range) to a virtual ground. The largest allowable current before overload
is around 3 µA (1 µA range) or 30 nA (10 nA range). No current larger than 10 mA
should ever be applied to this input.
Voltage [A
−
B]
The voltage input can be either a single-ended (A) or differential (A
−
B) voltage. The
shields of A and B are connected and grounded by either 10
Ω (Ground) or 10
kΩ (Float).
Voltage [Couple]
This key selects the voltage input coupling. The signal input can be either ac or dc
coupled. The ac coupling option uses a high pass filter to pass signals above 160 mHz
and attenuate signals at lower frequencies. ac coupling should be used at frequencies
above 160 mHz whenever possible. At lower frequencies, dc coupling is required. ac
Содержание SR860
Страница 1: ...Revision 2 01 Operation Manual SR860 500 kHz DSP Lock in Amplifier ...
Страница 5: ...Safety and Preparation For Use iii SR860 DSP Lock in Amplifier ...
Страница 6: ...iv Safety and Preparation For Use SR860 DSP Lock in Amplifier ...
Страница 54: ...36 Getting Started Chapter 1 SR860 DSP Lock in Amplifier ...
Страница 118: ......
Страница 172: ......
Страница 186: ...168 The FFT Display Appendix B SR860 DSP Lock in Amplifier ...
Страница 192: ......
Страница 222: ...204 Circuit Description Appendix H SR860 DSP Lock in Amplifier Partial schematics follow this page ...