Chapter 4
Programming
153
SR860 DSP Lock-in Amplifier
Lia Status Word
bit name
usage
0
CH1OV
CH1 output scale overload
1
CH2OV
CH2 output scale overload
2
unused
3
UNLK
External reference or Chop unlock detected.
4
RANGE
Input range overload detected
5
SYNCF
Sync filter frequency out of range
6
SYNCOV
Sync filter overload
7
TRIG
Set when data storage is triggered
8
DAT1OV
Data Channel 1 scale overload
9
DAT2OV
Data Channel 2 scale overload
10 DAT3OV
Data Channel 3 scale overload
11 DAT4OV
Data Channel 4 scale overload
12 DCAPFIN
Display capture to USB stick completed
13 SCNST
Scan started
14 SCNFIN
Scan completed
The LIA Status bits stay set until cleared by reading or by the
*CLS
command.
Error Status Byte
bit name
usage
0
CLK
External 10 MHz clock input error.
1
BACKUP
Battery backup failed.
2
unused
3
unused
4
VXI
VXI-11 error.
5
GPIB
GPIB fast data transfer mode aborted.
6
USBDEV
USB device error (interface error).
7
USBHOST
USB host error (memory stick error).
The Error Status bits stay set until cleared by reading or by the
*CLS
command.
Содержание SR860
Страница 1: ...Revision 2 01 Operation Manual SR860 500 kHz DSP Lock in Amplifier ...
Страница 5: ...Safety and Preparation For Use iii SR860 DSP Lock in Amplifier ...
Страница 6: ...iv Safety and Preparation For Use SR860 DSP Lock in Amplifier ...
Страница 54: ...36 Getting Started Chapter 1 SR860 DSP Lock in Amplifier ...
Страница 118: ......
Страница 172: ......
Страница 186: ...168 The FFT Display Appendix B SR860 DSP Lock in Amplifier ...
Страница 192: ......
Страница 222: ...204 Circuit Description Appendix H SR860 DSP Lock in Amplifier Partial schematics follow this page ...