Chapter 3
Operation
93
SR860 DSP Lock-in Amplifier
Ethernet Settings
Choose either Static or DHCP for the TCP/IP address mode. Highlight an address value
to change it using the numeric keypad.
GPIB and RS-232 Settings
Highlight the GPIB Address to change it using the numeric keypad.
Use the RS-232 Baud Rate {
↑} and {↓} buttons to select a baud rate. Select a terminator
and parity.
VXI-11 and Web Settings
The SR860 allows computers to connect via VXI-11 from a range of Trusted IP addresses
or Disabled altogether.
Set the Trusted IP address using ‘*’ as a wildcard. This allows any computer on a
specified subnet VXI-11 access.
The SR860 has a webserver which allows computer access thru a webpage (at the
SR860’s IP address). This access can be Enabled or Disabled. The webserver allows
instrument settings to be changed and the outputs to be monitored.