Commands
xv
SR860 DSP Lock-in Amplifier
SR860 Status Bytes
Serial Poll Status Byte
bit
name
usage
0
unused
1
unused
2
ERR
An enabled bit in the error status byte has been set
3
LIA
An enabled bit in the LIA status byte has been set
4
MAV
The interface output buffer is non-empty
5
ESB
An enabled bit in the standard status byte has been set
6
SRQ
SRQ (service request) has occurred
7
unused
Standard Event Status Byte
bit
name
usage
0
OPC
Operation complete
1
INP
Input queue overflow
2
unused
3
QRY
Output queue overflow
4
EXE
A command cannot execute correctly or a parameter is out of range
5
CMD
An illegal command is received
6
URQ
Set by any user front panel action
7
PON
Set by power-on
LIA Status Word
bit
name
usage
0
CH1OV
CH1 output overload
1
CH2OV
CH2 output overload
2
unused
3
UNLK
External reference or Chop unlock detected
4
RANGE
Input range overload detected
5
SYNCF
Sync filter frequency out of range
6
SYNCOV Sync filter overload
7
TRIG
Set when data storage is triggered
8
DAT1OV Data Channel 1 output overload
9
DAT2OV Data Channel 2 output overload
10
DAT3OV Data Channel 3 output overload
11
DAT4OV Data Channel 4 output overload
12
DCAPFIN Display capture to USB stick completed
13
SCNST
Scan started
14
SCNFIN Scan completed
Содержание SR860
Страница 1: ...Revision 2 01 Operation Manual SR860 500 kHz DSP Lock in Amplifier ...
Страница 5: ...Safety and Preparation For Use iii SR860 DSP Lock in Amplifier ...
Страница 6: ...iv Safety and Preparation For Use SR860 DSP Lock in Amplifier ...
Страница 54: ...36 Getting Started Chapter 1 SR860 DSP Lock in Amplifier ...
Страница 118: ......
Страница 172: ......
Страница 186: ...168 The FFT Display Appendix B SR860 DSP Lock in Amplifier ...
Страница 192: ......
Страница 222: ...204 Circuit Description Appendix H SR860 DSP Lock in Amplifier Partial schematics follow this page ...