Chapter 4
Programming
137
SR860 DSP Lock-in Amplifier
Note that when capturing one sample per trigger, this setting is ignored and one data
sample is always captured as soon as possible after detection of the trigger.
The
CAPTURERATE?
query returns the actual capture rate in Hz (not the value n).
Example
CAPTURERATE 4
Sets the capture rate to the max rate divided by 2
4
CAPTURERATE?
Returns the actual capture rate in Hz
CAPTURESTART [ ONEshot | CONTinuous | i ] , [ IMMediate | TRIGstart | SAMPpertrig | j ]
The
CAPTURESTART i, j
command starts data capture for either OneShot (i=0) or
Continiuous (i=1) acquisition. The parameter j selects whether capture starts immediately
(IMMediate start, j=0), is started or stopped by a hardware trigger (TRIGstart, j=1), or
captures one sample per hardware trigger (SAMPpertrig, j=2).
The hardware trigger is a TTL input on the rear panel. The active edge of the trigger is a
falling edge
. For hardware triggered start or stop, a triggered start requires using OneShot
acquisition, while a triggered stop required using Continuous acquisition.
Both parameters i and j are required in this command.
This command clears any previously captured data and starts a new capture.
Example
CAPTURESTART ONE, IMM
Starts OneShot capture immediately
CAPTURESTART 0, 0
CAPTURESTOP
The
CAPTURESTOP
command stops data capture in any mode.
If the capture is waiting for a hardware trigger to start, then the capture is aborted and the
trigger will be ignored.
If capture is in progress, then capture is halted. Any data already captured is preserved
and may be read. The remaining space in the current 2kB buffer is filled with zero.
If capture has already ended, the CAPTURESTOP command does nothing.
Example
CAPTURESTOP
Stop data capture at the next 2 kB boundary
CAPTURESTAT?
The
CAPTURESTAT?
query returns the data capture state. This is a query only
command.
The returned integer is a 3-bit binary encoded word
Bit
Weight Definition
0
1
Capture currently in progress
1
2
Capture triggered, either by
Содержание SR860
Страница 1: ...Revision 2 01 Operation Manual SR860 500 kHz DSP Lock in Amplifier ...
Страница 5: ...Safety and Preparation For Use iii SR860 DSP Lock in Amplifier ...
Страница 6: ...iv Safety and Preparation For Use SR860 DSP Lock in Amplifier ...
Страница 54: ...36 Getting Started Chapter 1 SR860 DSP Lock in Amplifier ...
Страница 118: ......
Страница 172: ......
Страница 186: ...168 The FFT Display Appendix B SR860 DSP Lock in Amplifier ...
Страница 192: ......
Страница 222: ...204 Circuit Description Appendix H SR860 DSP Lock in Amplifier Partial schematics follow this page ...