IM-P309-02
MI Issue 2
64
15.9.5 Status output simulation
The Status output simulation function allows the operator to purposefully activate the status
output.
Press
to display the current state.
Status output
(off)
As mentioned in Section 14.1.3.1 Selection window / make a selection, the operator can
toggle between "on" and "off".
15.9.6 Pulse output simulation
The Pulse output simulation function allows the operator to define a frequency to be assigned
to the pulse output. After selecting this function and pressing enter, the following selection
field will be display
Set frequency
0210.0 Hz
This field shows the current frequency. As mentioned in Section 14.1.3.2 "Input window /
modify a value", the definable frequency ranges from 6 Hz to 1100 Hz.
15.9.7 Current output simulation
This function allows the operator to define a current for current interface 1. Press
to display
the set current.
Set I1
I1 = 10.50 mA
As mentioned in Section 14.1.3.2 "Input window / modify a value", the current value can be
changed.