IM-P309-02
MI Issue 2
63
15.9.2 Simulation direct / preset value Q
This function allows the operator to define whether simulation is comprised of the measurement
of the volume flow or whether the outputs will be set directly. Press
to display the selected
type of simulation.
Simulation
[direct]
As mentioned in Section 14.1.3.1 Selection window / make a selection, the operator can
choose between the following settings:
-
Direct
pulse and current outputs are programmed directly
-
QVabs
a measurement is simulated
If "direct" simulation is activated, any output will perform based on the settings described in
Sections 15.9.5 Status output simulation to 15.9.5.7 Current output simulation. It is therefore
recommended that the settings be defined before starting simulation. They can then be
purposefully changed during simulation.
Simulation will be deactivated automatically if the operator switched the device
off or did not touch any control unit keys for about 10 minutes.
15.9.3 Simulation measured flow Q
If the operator selected the setting "QVabs" described in Section 15.9.2, the following settings
of a volume flow will affect the output behavior during measured value simulation.
In order to simulate volume flow, the operator can define a "measured value." The flow rates
will be simulated in both directions. All outputs will perform based on the simulated measured
value.
Preset QVabs
±0900.0 l/h
The simulation value is entered as described in Section 14.1.3.2 "Input window / modify a
value".
15.9.4 Direct simulation of outputs
If the operator selected the setting "Direct simulation" described in Section 15.9.2 „Simulation
direct ", the following 3 possible settings will affect the output. All outputs are simulated at
the same time by these settings.