![Spacelabs Healthcare Ultraview 90341 Скачать руководство пользователя страница 191](http://html1.mh-extra.com/html/spacelabs-healthcare/ultraview-90341/ultraview-90341_service-manual_1339025191.webp)
www.spacelabshealthcare.com
A - 1
U
L T R A V I E W
D
I G I T A L
T
E L E M E T R Y
- S
E R V I C E
M
A N U A L
Appendix A — Electromagnetic
Compatibility
Electromagnetic Emissions
Note:
This equipment has been tested under laboratory conditions and
is suitable for use in all establishments, including domestic
buildings and other buildings directly connected to the public
low-voltage power supply network. The customer or user of this
equipment should ensure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic Environment
RF emissions
CISPR 22
Class B
This equipment uses RF energy only for internal
function. Therefore, RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment
Harmonic emissions
IEC 61000-3-2
Complies
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage fluctuations/
flicker